AC-5200 Mini UV Monitor > UV Monitor, Fraction Collector

본문 바로가기

쇼핑몰 검색

Home AC-5200 Mini UV Monitor > UV Monitor, Fraction Collector
  • 제품소개 
  • 1D Gel Electrophoresis 2D Gel Electrophoresis Agarose Gel Electrophoresis Precast Gel Hand Gel Casting Power Supply Electrophoresis Reagents Western Blotting Blotting Reagents & Membranes Heat Block, Shaker Gel Documentation Luminescent Detector Luminescent Reagents Perista® Pump UV Monitor, Fraction Collector Reagents Live Cell Imaging

    AC-5200 Mini UV Monitor > UV Monitor, Fraction Collector

    AC-5200 Mini UV Monitor

    상품 상세설명

    UV monitor with digital display to measure UV absorption of liquid eluting from chromatography system. 

    AC-5200 Mini UV Monitor

    AC-5200 Mini UV Monitor

    Purpose and Application

    AC-5200 Mini UV Monitor measures UV absorption of eluate from liquid chromatography system. 

    Features
    • Digital displays of absorbance and absorbance range are equipped.
    • Recorder output (0 - 10mV) and computer output (1V/Abs) are provided.
    • Optical system is compact and minimizes dead volume.
    • Flow cell is easy to remove and maintain, and can be autoclaved.
    • Various flow cells with different light path length and volume are available.
    • Various wavelength filters are available.
    Specifications
    Product NumberAC-5200
      AC-5200S (Optical Device) AC-5200L (Control Device)
    Product NameMini UV Monitor
    Light sourceLow-pressure mercury lamp
    Optical DeviceSilicone Photo Diode
    Wavelength254 nm, 280 nm (2 wavelength filters are included standard accessories)
    Absorbance Range 0.005/ 0.01/ 0.02/ 0.05/ 0.1/ 0.2/ 0.5/ 1.0/ 2.0 Abs
     (9 steps selectable)
    DisplayAbsorbance, Absorbance Range
    Event MarkerAbout - 1.5mV, MARK Switch ON Time, 
     External Marker Control, Make Contact Input
    Balance AdjustmentAuto Zero: Range (from +2 to -0.5 Abs), 
     External Control (Over 100 ms: Make Contact)
    Flow CellQuartz, Optical Path Length (5 mm), Volume (60 μL), Inner Diameter φ1.7 - 0.9 mm Teflon Tube
    Absorbance OutputRecorder (0 - 10 mV), Computer (1 V/Abs) 
    Environmental ConditionRoom Temperature 2 - 35 °C / Relative Humidity should be under 90 %
    PowerAC100±10V  50/60Hz  8VA
    DimentionsAC-5200S (Optical Device): 69(W)x144(D)x74(H)mm
    AC-5200L (Control Device): 164(W)x184(D)×74(H)mm
    WeightAC-5200S (Optical Device): 0.8 kg
    AC-5200L (Control Device): 1.8 kg

     

    상품문의

    상품문의가 없습니다.

    상단